Panther Fusion GBS Assay REF PRD-04484 Kit containing Panther Fusion GBS Reagent Cartridge IVD.

Company
Hologic, Inc
Recall Initiated
December 13, 2024
Posted
January 14, 2025
Recall Number
Z-0907-2025
Quantity
380
Firm Location
San Diego, CA

Reason for Recall

Potential weak pouch seal on Panther Fusion GBS assay cartridge, causing potential for invalid or incorrect results leading to a possible delay of treatment.

Distribution

AR, CA, CT, FL, IN, KS, LA, MA, ME, MI, NC, NJ, NV, NY, PR, TX, UT, VT, WA, WV, and Canada.

Lot / Code Info

UDI-DI: 15420045510890, Kit Lot 715927, containing Cartridge Lot 624907, Exp. Date 2025-08-15.

Root Cause

Process control

Action Taken

An Urgent Medical Device Recall Notification letter issued 12/13/2024 via Fed Ex. The letter instructs customers to discontinue use of the Panther Fusion GBS Assay kit lot 715927 containing cartridge lot 624907 and segregate and count the number of affected Panther Fusion GBS Assay kits from lot 715927 containing cartridge lot 624907. Customers are instructed to destroy the product and fill out the online Customer Response Form (CRF) indicating that the on-hand inventory of recalled Panther Fusion GBS Assay Kit has been destroyed. If you have any questions or concerns regarding this recall notification, please contact Hologic Technical Support at +1 888 484 4747 or +1 858 410 8511 during the hours of 5 am PT and 5 pm PT, or by e-mail at molecularsupport@hologic.com.