Medfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Company
Smiths Medical Asd Inc.
Recall Initiated
June 26, 2020
Posted
August 6, 2020
Recall Number
Z-2734-2020
Quantity
35616 units
Firm Location
Minneapolis, MN

Reason for Recall

Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.

Distribution

Worldwide distribution - United Arab Emirates, United States, Afghanistan, Albania, Australia, Belgium, Bermuda, Canada, China, Germany, Denmark, Great Britain, Hong Kong, Indonesia, Ireland, India, Korea, Kuwait, Lebanon, Mexico, Malaysia, New Zealand, Oman, Panama, Philippines, Pakistan, Poland, Saudi Arabia, Trinidad and Tobago

Lot / Code Info

Model Numbers: 4000-0105-249, 4000-0105-50, 4000-0105-51, 4000-0105-78, 4000-0106-00, 4000-0106-01, and 4000-0106-231. Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1

Root Cause

Software design

Action Taken

Consignees are being contacted via direct mailing and email. Consignees are asked to review and complete the Recall Response form indicating understanding of notification. Distributors are asked to forward the notification to parties to whom product has been distributed. Smiths Medical will send out 2 additional notifications by email. Phone calls will be made to non-responders.