Beacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.

Company
Covidien Llc
Recall Initiated
June 26, 2020
Posted
August 5, 2020
Recall Number
Z-2769-2020
Quantity
269 units
Firm Location
Mansfield, MA

Reason for Recall

This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.

Distribution

world wide distribution

Lot / Code Info

Model # DSA-135-01 UDI: 10884521663527 Lot #: F2510371X, F2510861X, F2512341X, F2512383X, F2512493X, F2513453X

Root Cause

Under Investigation by firm

Action Taken

Medtronic sent a letter notification to customers dated July 06, 2020 recalling effected lots of product. This letter instructed customers to discontinue use of the product and quarantine effected inventory. Customers were also instructed to return effected product and complete recall response forms accordingly.