RecallDepth

RP500e Handheld Barcode Scanner, Zebra Model: DS4308 - HC0062BZZWW, Siemens Material Number (SMN) 11416778

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
May 27, 2020
Terminated
January 12, 2021
Recall Number
Z-2432-2020
Quantity
436
Firm Location
Norwood, MA

Reason for Recall

The scanner is not confirming the integrity of read barcode data using the check-digit when reading Code 39 barcodes.

Distribution

Foreign distribution only.

Lot / Code Info

All units

Root Cause

Software design

Action Taken

On May 27, 2020, the firm notified customers via Urgent Field Safety Notice letters. Product was distributed to foreign customers only. Customers were instructed to complete and return the included customer response form. Customers whose systems are not configured to use code 39 with check-digit need take no further action. However, for customers whose systems are configured to use code 39 with check-digit, the customer notice provided two workaround options: 1) Use the integrated on-board barcode scanner, or 2) Re-configure the handheld barcode scanner following the instructions in the letter. If you have any questions, please contact your Siemens Customer Care Center or your local Siemens technical support representative

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →