RecallDepth

Stratus CS Acute Care cTNI TestPak.IVD test for cardiac troponin I in heparinized plasma.

Company
Siemens Healthcare Diagnostics Inc
Recall Initiated
March 31, 2025
Posted
April 25, 2025
Recall Number
Z-1673-2025
Quantity
367 units
Firm Location
Norwood, MA

Reason for Recall

Siemens Healthcare Diagnostics Inc., is recalling their Stratus CS Acute Care cTNI TestPak, an IVD test for cardiac troponin I in heparinized plasma by removal. The reason for the recall is an increased occurrence of random non-repeatable false positive cardiac Troponin l(cTnI) results at any point during the TestPaks shelf life when using the Stratus CS cTnI Acute Care Testpak.

Distribution

Worldwide - US Nationwide distribution in the states of CA, IL, KS, MI, MO, NJ, NY, OH, OK, PA, SC & WI. The countries of Austria, Canada, Chile, France, Germany, Iraq, Italy, Malaysia, North Macedonia, Saudi Arabia, Serbia, Slovenia, Spain, Sweden, Switzerland & Vietnam.

Lot / Code Info

Siemens Materials No. 10445071; UDI-DI: 0405686902174VK; Lot No. 234337002 and forward.

Root Cause

Under Investigation by firm

Action Taken

An Urgent Medical Device Correction notice dated March 2025 was mailed to consignees. The notice instructed consignees to review the provided notification with their medical director to determine the appropriate course of action. Siemens will reimburse users for repeat tests and discarded units associated with this recall; consignees should contact the Customer Care Center for reimbursement. If product was further distributed, Siemens asks that the recall notification be forwarded. The recall notification is to be retained with laboratory records and forwarded to those who use the recalled devices. Consignees with any questions can contact Siemens Healthineers Technical Solutions Center at 977-229-3711.

More recalls by Siemens Healthcare Diagnostics Inc

View all recalls by this company →