RecallDepth

ImplantDirect Legacy(TM) 3 Implant, 13MM L, REF 864713 This Legacy implant is a two-piece implant for one-stage or two-stage surgical procedures. These implants are intended for use in partially and fully edentulous upper and lower jaws in support of single or multiple-unit restorations and terminal or intermediate abutment support for fixed bridgework.

Company
Implant Direct Sybron Manufacturing, Llc
Recall Initiated
October 10, 2017
Terminated
September 27, 2018
Recall Number
Z-0185-2018
Quantity
75 units
Firm Location
Westlake Village, CA

Reason for Recall

One reported complaint and discovered that the outer vial was out of round and undersized. Therefore a portion of the opening may not allow inner vial and components to slide freely out of the vial. As a result, the cap could possibly not be screwed on properly and the sterilization compromised.

Distribution

Worldwide distribution: US (nationwide) including Puerto Rico and country of Japan.

Lot / Code Info

Lot Number: 79607

Root Cause

Process design

Action Taken

The firm, Implant Direct Sybron Manufacturing, LLC, sent an "URGENT: Safety Alert/Advisory Notice" letter to its distributors/customers on 10/10/2017. Distributors were instructed as follows: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form within 48 hours. 3. If you are an authorized Implant Direct Sybron Manufacturing distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within fortyeight (48) hours of receipt of this notification in order to provide the customers with replacement product." Enclosed within this notification, you will find a padded return envelope with replacement product. Please return affected product in the padded envelope. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. Customers were instructed as follows: "1. Please review your inventory for the affected product. 2. Please complete and return the Acknowledgement and Recall Return Form within 48 hours. Please share this information with your staff and retain this notification as part of your documentation. Enclosed within this notification, you will find a padded return envelope with replacement product. Please return affected product in the padded envelope. If you have any questions contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425."

More recalls by Implant Direct Sybron Manufacturing, Llc

View all recalls by this company →