RecallDepth

ImplantDirect, REF 834713, Legacy 2 Implant 4.7mmD x 13mmL, 4.5mmD Platform,

Company
Implant Direct Sybron Manufacturing, Llc
Recall Initiated
September 23, 2021
Recall Number
Z-0198-2022
Quantity
83 implants
Firm Location
Westlake Village, CA

Reason for Recall

Packaged dental implant contains a different size then the size declared on the labeling.

Distribution

Worldwide distribution - US Nationwide distribution in the states of AZ, CA, HI, IL, KS, LA, MA, MD, MI, MO, MT, NE, NJ, NY, OH, OR, PA, RI, SC, TX, UT, VA, WA, WI and the country of Japan.

Lot / Code Info

lot # 166300 / UDI: 10841307101796

Root Cause

Packaging change control

Action Taken

On October 11, 2021, Implant Direct issued an "Urgent: Medical Device Recall" Notice via. Mail . In addition to informing consignees about the recall, the firm ask consignees to take the following actions: 1. Please immediately examine, identify, and quarantine the affected product subject to recall (Table-1). 2. If located, immediately return the affected product with the pre-paid shipping label included with the notification letter to us according to the instructions on the attached form. Upon receipt of the returned product a credit will be issued accordingly. 3. Please complete, scan, and email the Acknowledgement Form within forty eight (48) hours of receipt of this notification, whether you have affected product or not, to legacy2implants@implantdirect.com 4. If you are an authorized Implant Direct Sybron Manufacturing LLC distributor, we request that you identify those customers that may have been shipped the affected product lot and contact these customers to inform them of this issue within forty eight (48) hours of receipt of this notification. 5. If you have any questions, contact Implant Direct Sybron Manufacturing LLC Customer Care at 1-888-649-6425. If you or your patients have experienced any issues as a result of the affected products noted in this communication, you may voluntarily report the incident to the FDA through the MEDWATCH reporting system at the following: http://www.fda.gov/Safety/MedWatch/HowToReport/ucm053074.htm Implant Direct Sybron Manufacturing sincerely apologizes for the inconvenience this situation may cause. 6. For Questions Contact: Customer Service and Technical Support team Monday through Friday, 8:00 AM to 4:30 PM, Pacific Time Zone Tel: 888-649-6425 Email: legacy2implants@implantdirect.com

More recalls by Implant Direct Sybron Manufacturing, Llc

View all recalls by this company →