RecallDepth

Peri-Loc(R) VLP 3.5 MM X 6 MM SELF-TAPPING CORTEX SCREW, QTY: (1), STERILE R, REF 71801306 Orthopedic

This recall has been terminated (originally issued March 9, 2017).

Company
Smith & Nephew Inc
Recall Initiated
March 9, 2017
Terminated
August 21, 2019
Recall Number
Z-1854-2017
Quantity
1 unit
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

The expiration dates on the labeling are incorrect, the actual expiration dates are before the date on the labeling.

Distribution

Worldwide Distribution - US (Nationwide) Internationally to Colombia, Denmark, Finland, France, Great Britain, Italy, and Singapore

Lot / Code Info

Batch 15HM05197

Root Cause

Labeling Change Control

Action Taken

Smith & Nephew sent an Urgent Medical First Recall Notice dated March 9, 2017, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The letter requested the quarantine and return of the devices, and for the distributors to notify their consignees and ensure the same actions. International consignees were notified by email. Consignees with questions were instructed to contact FieldAction@smith-nephew.com. For questions regarding this recall call 978-749-1440.