RecallDepth

smith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)

This recall has been terminated (originally issued May 4, 2018).

Company
Smith & Nephew Inc
Recall Initiated
May 4, 2018
Terminated
February 22, 2019
Recall Number
Z-2163-2018
Quantity
5 units
Firm Location
Memphis, TN
Official Source
View on FDA website ↗

Reason for Recall

One lot of GENESIS II 13 mm tibial punches are 2 mm smaller than specification.

Distribution

International only: United Arab Emirates, Spain, Mexico

Lot / Code Info

15FM16181

Root Cause

Error in labeling

Action Taken

The firm initiated the recall by email on 05/04/2018. Consignees are asked to inspect inventory, locate and quarantine affected, unused devices. Affected product will be returned to Smith & Nephew. All consignees are asked to complete the Response Form, even if they do not have product to return.

More recalls by Smith & Nephew Inc

View all recalls by this company →