iConnect Access software. The firm name on the label is Merge Healthcare, Hartland, WI.

This recall has been terminated (originally issued May 25, 2016).

Company
Merge Healthcare, Inc.
Recall Initiated
May 25, 2016
Terminated
August 28, 2019
Recall Number
Z-0611-2017
Quantity
60 sites potentially have the affected versions
Firm Location
Hartland, WI
Official Source
View on FDA website ↗

Reason for Recall

Software displayed incorrect prior reports in the viewport area, only when more than one prior study (2 or more) was viewed.

Distribution

Distribution was made to medical facilities nationwide. Foreign distribution was also made.

Lot / Code Info

Versions 3.0.1, 3.1, and 3.2

Root Cause

Software design

Action Taken

The recalling firm issued a notification letter dated 1/27/2016 via email or via certified mail if the firm did not have an email address for the customer. Nonresponders were issued a second letter dated 5/25/2016 on 5/26/2016.