Instructions for Use (IFU) for Progesterone 3 Calibrator Set, repackaged as one set/package. The Instructions for Use (IFU) included in the package was for the incorrect lot number. Only the IFU required correction.

This recall has been terminated (originally issued July 25, 2016).

Company
Tosoh Bioscience, Inc.
Recall Initiated
June 7, 2016
Posted
July 25, 2016
Terminated
February 22, 2017
Recall Number
Z-2282-2016
Quantity
41 single sets of Progesterone 3 calibrators
Firm Location
South San Francisco, CA
Official Source
View on FDA website ↗

Reason for Recall

Instructions For Use had incorrect lot number and calibrator values

Distribution

US Nationwide Distribution in the states of AL; CA; CO; FL; KY; MA; NY; OH; SC; TX; UT; VA and WI.

Lot / Code Info

Catalog Number 02534001 (One set of Calibrators US)  Lot G13C303 (only those received/repackaged in the US)

Root Cause

Error in labeling

Action Taken

Tosoh Bioscience sent an Urgent Correction Notification letter dated June 7, 2016 to all affected customers. The letter identified the affected product, problem and actions to be taken. For questions call Tosoh Technical Support at 1-800-248-6764.