AIA-PACK TgAb CONTROL SET

This recall has been terminated (originally issued May 15, 2017).

Company
Tosoh Bioscience, Inc.
Recall Initiated
February 8, 2017
Posted
May 15, 2017
Terminated
February 8, 2019
Recall Number
Z-2169-2017
Quantity
64 boxes
Firm Location
South San Francisco, CA
Official Source
View on FDA website ↗

Reason for Recall

The label contains two different methodologies. Quality Control values for non-FDA approved methodology (CL) are on the label in addition to the FDA approved methodology values (AIA). Two sets of instructions were included on the label.

Distribution

Product was shipped to the following states: AK, FL, GA, MA, NY, NY, OH, OK & TX.

Lot / Code Info

TgAb: G644899 Expires June 2017; G844801 Expires August 2017 & GX44802 Expires 09/30/2017.

Root Cause

Labeling mix-ups

Action Taken

On 3/3/2017 the firm sent recall notificaion letters to their customers.