AIA-PACK TgAb CONTROL SET
This recall has been terminated (originally issued May 15, 2017).
- Company
- Tosoh Bioscience, Inc.
- Recall Initiated
- February 8, 2017
- Posted
- May 15, 2017
- Terminated
- February 8, 2019
- Recall Number
- Z-2169-2017
- Quantity
- 64 boxes
- Firm Location
- South San Francisco, CA
- Official Source
- View on FDA website ↗
Reason for Recall
The label contains two different methodologies. Quality Control values for non-FDA approved methodology (CL) are on the label in addition to the FDA approved methodology values (AIA). Two sets of instructions were included on the label.
Distribution
Product was shipped to the following states: AK, FL, GA, MA, NY, NY, OH, OK & TX.
Lot / Code Info
TgAb: G644899 Expires June 2017; G844801 Expires August 2017 & GX44802 Expires 09/30/2017.
Root Cause
Labeling mix-ups
Action Taken
On 3/3/2017 the firm sent recall notificaion letters to their customers.
More recalls by Tosoh Bioscience, Inc.
AIA-PACK TPOAb CONTROL SET
May 15, 2017
Instructions for Use (IFU) for Progesterone 3 Calibrator Set, repackaged as one set/package.
The...
Jul 25, 2016