FUJIFILM Synapse PACS Software: versions 7.0.0 and 7.0.1 - Product Usage: intended for use as a web based application on an off-the shelf PC which meets or exceeds minimum specifications and is networked with a FUJIFILM Synapse PACS server.

Company
Fujifilm Medical Systems U.S.A., Inc.
Recall Initiated
December 11, 2020
Terminated
November 17, 2021
Recall Number
Z-0918-2021
Quantity
1 unit
Firm Location
Lexington, MA

Reason for Recall

The software does not update measurements and calculations in the Clinical Reporting Application (CRA) when the ventricular trace is changed in the study by a different user.

Distribution

US Nationwide distribution in the state of MD.

Lot / Code Info

Versions 7.0.0 and 7.0.1

Root Cause

Software change control

Action Taken

On 12/11/2020 Fujifilm sent consignee notifications via FedEx Priority Overnight stating that there are issues with Synapse PACS Software Version 7.0.000 and 7.0.100. Fujifilm provided a work around until a hotfix is scheduled and completed.