TASWako Chip Cassette is part of the Wako TASWako i30 kit
- Recall Initiated
- February 18, 2021
- Terminated
- February 24, 2022
- Recall Number
- Z-1412-2021
- Quantity
- 496 units
- Firm Location
- Lexington, MA
Reason for Recall
Potentially defective Chip Cassette units, when used for DCP or AFP-L3% measurement, incorrect DCP and AFP-L3% values may result
Distribution
CA, LA, MN, NC, UT, WI
Lot / Code Info
Lot Number: EQ194 Exp. Date: 05/31/2022
Root Cause
Under Investigation by firm
Action Taken
Fujifilm issued Consignee notifications via courier delivery and electronic mail on February 18, 2021. The letter states reason for recall, health risk and action to take: FUJIFILM is recommending that you immediately cease use of any Cavity #2 chips in Lot EQ194 and return the product to FUJIFILM. Please complete and return the Field Action Verification Form attached on page 3 of this communication. Providing the response with the information requested is essential for ensuring appropriate action is taken: Upon receipt of the Field Action Verification Form, FUJIFILM will provide replacement chip cassette boxes (each containing 5 x 20 chips) from a different lot. Please contact our field action coordinator, Jeffrey Wan, via phone at 201-675-8947 or email at jeffrey.wan@fujifilm.com, if you have any further questions regarding this field action.