REF 0930 Adult Elongated CO2/O2 Mask, 7' Tubing w/ Threaded Nut, Sivel GSE, 10' GSL X .038" ID X .107" OD, Reflective Luer Connectors, Rx Only, CE0482, (01)00709078005718 - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.

This recall has been terminated (originally issued November 11, 2020).

Company
Westmed Inc
Recall Initiated
November 11, 2020
Terminated
February 7, 2023
Recall Number
Z-0743-2021
Quantity
600 units
Firm Location
Tucson, AZ
Official Source
View on FDA website ↗

Reason for Recall

There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom

Lot / Code Info

Lot # 091620N04

Root Cause

Process control

Action Taken

On November 11, 2020, West Med, Inc. issued a recall notification for oxygen delivery products via E-Mail to affected consignees. In addition to informing consignees about the recall products, the firm asked consignees to take the following actions: 1. Please complete the attached response form and quarantine affected product. 2. DO NOT DESTROY OR DISPOSE OF PRODUCT AT THIS TIME. 3. Contact Kalyn Galvez-Latneau, Westmed Customer Service Manager, for additional information and to obtain a Return Material Authorization (RMA). Kalyn Galvez-Latneau, kgalvez@westmedinc.com, (520) 294-7987, Ext 1232.