REF 0562K (Adult Elongated O2/CO2 Mask - Product Usage: intended to connect with an oxygen source to deliver oxygen therapy to the patient.

Company
Westmed Inc
Recall Initiated
November 11, 2020
Terminated
February 7, 2023
Recall Number
Z-0745-2021
Quantity
16,000 units
Firm Location
Tucson, AZ

Reason for Recall

There is a potential that the fitment between the oxygen delivery tube and the mask adapter may become loose. This could result in the device tubing not making a secure connection to the mask adapter resulting in a possible delay in treatment.

Distribution

Worldwide distribution - US Nationwide distribution including in the states of PA, OK, MO, CO, RI, IL, CA, TX, OH, MD, IL, MN, KS and the countries of Canada, Korea, United Kingdom

Lot / Code Info

Lot # 081920N38 (Distributed outside US)

Root Cause

Process control

Action Taken

On November 11, 2020, West Med, Inc. issued a recall notification for oxygen delivery products via E-Mail to affected consignees. In addition to informing consignees about the recall products, the firm asked consignees to take the following actions: 1. Please complete the attached response form and quarantine affected product. 2. DO NOT DESTROY OR DISPOSE OF PRODUCT AT THIS TIME. 3. Contact Kalyn Galvez-Latneau, Westmed Customer Service Manager, for additional information and to obtain a Return Material Authorization (RMA). Kalyn Galvez-Latneau, kgalvez@westmedinc.com, (520) 294-7987, Ext 1232.