OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.

This recall has been terminated (originally issued August 27, 2015).

Company
Nidek Inc
Recall Initiated
June 15, 2015
Posted
August 27, 2015
Terminated
April 25, 2016
Recall Number
Z-2477-2015
Quantity
3836
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.

Distribution

Nationwide Distribution.

Lot / Code Info

Software versions: 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01.

Root Cause

Software design

Action Taken

Dear Doctor letters mailing started on July 15, 2015 - first 100. They will be mailed out as the correction is received from Nidek, Japan. Dear Doctor letter will include software upgrade kit including CD or USB, instruction for upgrade and pre-paid response card to be returned to Nidek Inc.