OPD-Scan III Refractive Power/Corneal Analyzer Opthalmic Software versions 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01. Opthalmic: The OPD-Scan III is a diagnostic instrument that is indicated for use for Mapping of refractive error distribution of the eye by measurement and analysis of spherical power, cylindrical power, and cylinder axis.
This recall has been terminated (originally issued August 27, 2015).
- Company
- Nidek Inc
- Recall Initiated
- June 15, 2015
- Posted
- August 27, 2015
- Terminated
- April 25, 2016
- Recall Number
- Z-2477-2015
- Quantity
- 3836
- Firm Location
- Fremont, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Software bug was found where there was no difference in Total and Corneal high-order aberrations, but differences were found in Internal high-order aberrations.
Distribution
Nationwide Distribution.
Lot / Code Info
Software versions: 1.00.08 1.05.07 1.10.01, 1.01.02 1.06.02 1.11.02, 1.02.01 1.07.01 1.12.03, 1.03.02 1.08.01 1.13.01, 1.04.03 1.09.01.
Root Cause
Software design
Action Taken
Dear Doctor letters mailing started on July 15, 2015 - first 100. They will be mailed out as the correction is received from Nidek, Japan. Dear Doctor letter will include software upgrade kit including CD or USB, instruction for upgrade and pre-paid response card to be returned to Nidek Inc.