SPECULAR MICROSCOPE CEM-530; Software version 1.08 and 1.09. Opthalmic: The NIDEK Specular Microscope CEM-530 provides non-contact, high magnification image capture of endothelium enabling observation of the size and shape of cells. Information such as the number of endothelial cells, cell density, and cell area is analyzed through the captured images. The captured images and analysis results of endothelium are used for intraocular or corneal surgery, postoperative follow-up and corneal observation such as for endothelial disorders or the corneal state of patients who wear extended-wear contact lenses.

This recall has been terminated (originally issued September 1, 2016).

Company
Nidek Inc
Recall Initiated
August 8, 2016
Posted
September 1, 2016
Terminated
January 9, 2017
Recall Number
Z-2711-2016
Quantity
29 affected devices
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA.

Distribution

US Distribution to states of: TX, KS, FL, NV, CA, PA, KY, OR, TN, AL, VA, MO, OK, CT, NY. LA, NM, and IN.

Lot / Code Info

Microscopes with serial numbers: 120004,120005, 120006, 120007, 120009, 120010, 120011,120012, 120013, 120014, 0120016, 120019, 120021, 120023, 120024, 120026, 120030, 120036, 0120038, 120039, 120040, 120042, 120044,120045, 120046, 120047, 120048, 120049, 120050.

Root Cause

Software change control

Action Taken

The firm, Nidek, sent a Dear Doctor letters entitled "RECALL NOTIFICATION" dated August 15, 2016 to affected sites by Federal Express. The letter informs users of the issue and stated that Nidek field service representatives will visit to install updated software. If you have any questions, please call Customer Service at 1-800-722-0219; Monday-Friday 8am to 5pm PST or email: cem530update@noritsuservice.com.