Mevatron MD (MD, MDX); Siemens Healthcare Concord, CA 94520 Product Usage: PRIMUS and MEVATRON Linear Accelerators are intended to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer.
This recall has been terminated (originally issued January 17, 2012).
- Recall Initiated
- December 1, 2011
- Posted
- January 17, 2012
- Terminated
- December 2, 2013
- Recall Number
- Z-0789-2012
- Quantity
- 327 Active Devices - all models
- Firm Location
- Concord, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Bolts holding the trim weight may break causing the plates of the counterweight to become loose and fall into the cover.
Distribution
Worldwide Distribution - USA (nationwide)
Lot / Code Info
Part number 8319808, all units
Root Cause
Nonconforming Material/Component
Action Taken
Siemens sent a Customer Information Letter on December 1, 2011 to all affected customers. The letter idenfied what is the issue, when does the issue occur and what Siemens will do to address the issue. The letter recommend that the Customer Information letter be included in the Linear Accelerator System Owner manual chapter Safety Advisory Letters.