Baxter Colleague CX Infusion Pump, Single Channel Volumetric Infusion Pumps. Product Code: 2M8161.

This recall has been terminated (originally issued January 11, 2012).

Company
Baxter Healthcare Corp. Rt
Recall Initiated
May 8, 2009
Posted
January 11, 2012
Terminated
January 3, 2012
Recall Number
Z-0598-2012
Quantity
one pump
Firm Location
Round Lake, IL
Official Source
View on FDA website ↗

Reason for Recall

There was no documentation to show the required tests were performed following the replacement of the battery and battery harness.

Distribution

Puerto Rico

Lot / Code Info

Product Code: 2M8161, Serial Number 12050019CC

Root Cause

Process control

Action Taken

Baxter sent a service technician to the consignee location on 05/08/2009 to perform the battery check and discharge test. The pump was found to be within specification.