Animas 2020 Insulin Infusion Pump. Animas Corp. An insulin infusion.

This recall has been terminated (originally issued August 12, 2011).

Company
Animas Corporation
Recall Initiated
July 29, 2011
Posted
August 12, 2011
Terminated
December 7, 2011
Recall Number
Z-3017-2011
Quantity
380
Firm Location
West Chester, PA
Official Source
View on FDA website ↗

Reason for Recall

Incorrect component (1nF capacitor) was used to produce 191 printed circuit boards used in six pump lots of which 380 pumps were distributed. This issue may result in the pump displaying a "Replace Battery" alarm not related to the condition of the battery.

Distribution

Worldwide Distribution: Nationwide (USA) distribution including the states of CA, CO, MA, and MD; and the countries of Australia, Finland, France, Israel, Italy, Sweden and the U.K.

Lot / Code Info

Animas 2020 Insulin Pump  Serial numbers 00-89133-12, 00-89207-12, 14-88703-12, 14-88710-12, 15-88704-12, 15-89190-12, 16-89096-12, 16-89191-12, 17-88720-12, 17-89097-12, 17-89185-12, 17-89611-12, 18-88721-12, 18-89612-12, 19-88715, 18-89099-12, 19-89187-12, 19-89194-12, 19-89620-12, 20-88716-12, 20-89188-12, 20-89621-12, 21-88717-12, 21-88731-12, 21-89196-12, 21-89615-12, 22-88732-12, 22-89616-12, 23-88733-12, 24-88734-12, 25-88728-12, 25-88735-12, 27-88737-12, 28-88738-12, 33-89824-12, 35-89819-12, 36-89827-12, 88-89100-12, 91-89110-12, 92-89111-12, 93-89112-12, 94-89106-12, 94-89120-12, 95-89107-12, 95-89114-12, 95-89121-12, 96-89115-12, 21-88386-12, 21-88393-12, 22-88387-12, 96-89122-12, 96-89210-12, 97-89116-12, 97-89123-12, 97-89130-12, 01-89215-12, 02-89216-12, 02-89223-12, 02-89230-12, 03-89217-12, 03-89224-12, 03-89400-12, 04-89218-12, 04-89225-12, 04-89232-12, 04-89401-12, 05-89219-12, 05-89226-12, 06-89227-12, 06-89234-12, 06-89241-12, 06-89410-12, 07-89235-12, 07-89242-12, 07-89404-12, 07-89411-12, 08-89236-12, 08-89412-12, 09-89237-12, 09-89406-12, 09-89413-12, 10-89238-12, 10-89414-12, 11-89239-12, 11-89408-12, 11-89415-12, 12-89409-12, 13-89417-12, 14-89418-12, 14-89520-12, 15-89521-12, 17-89509-12, 17-89516-12, 18-89517-12, 18-98524-12, 18-89531-12, 19-89525-12, 29-88672-12, 32-88668-12, 32-88682-12, 20-89533-12, 20-89540-12, 21-88643-12, 21-89541-12, 22-89528-12, 22-89535-12, 22-89542-12, 23-88638-12, 23-88740-12, 24-88741-12, 24-89544-12, 25-88647-12, 26-88655-12, 26-89553-12, 27-88751-12, 27-89385-12, 27-89554-12, 27-89561-12, 28-88745-12, 28-89555-12, 29-88760-12, 29-89556-12, 29-89563-12, 30-88747-12, 30-88754-12, 30-88761-12, 30-89395-12, 31-88755-12, 31-88762-12, 31-89558-12, 32-88756-12, 32-89397-12, 33-88757-12, 33-88771-12, 34-88758-12, 34-88772-12, 35-88759-12, 35-88766-12, 36-88774-12, 37-88768-12, 39-88777-12, 39-88939-12, 41-88948-12, 42-88787-12, 51-88979-12, 03-88414-12, 10-89245-12, 33-88690-12, 40-88697-12, 98-89131-12, 14-89256-12, 15-89257-12, 16-89258-12, 17-89273-12, 18-89267-12, 19-89444-12, 20-89269-12, 21-89460-12, 22-88901-12, 22-89461-12, 24-89463-12, 26-88905-12, 27-89466-12, 33-88852-12, 34-88853-12, 35-88854-12, 36-88855-12, 37-88856-12, 39-88872-12, 41-88874-12, 43-88869-12, 47-88887-12, 99-88403-12, 00-88404-12, 00-88411-12, 01-88412-12, 02-88406-12, 02-88413-12, 03-88407-12, 04-88408-12, 04-88415-12, 12-88370-12, 13-88371-12, 14-88372-12, 15-88373-12, 16-88367-12, 16-88381-12, 17-88375-12, 18-88376-12, 18-88383-12, 18-88390-12, 19-88377-12, 19-88384-12, 19-88391-12, 20-88378-12, 20-88385-12, 20-88392-12, 44-88789-12, 00-89140-12, 01-89141-12, 64-83146-12, 00-89221-12, 05-89240-12, 17-89266-12, 22-88394-12, 23-88388-12, 23-88395-12, 24-88389-12, 24-88396-12, 25-88397-12, 26-88398-12, 27-88399-12, 96-88400-12, 99-88410-12, 38-88695-12, 41-88698-12, 09-88603-12, 10-88604-12, 12-88606-12, 12-88613-12, 13-88614-12, 15-88616-12, 22-88637-12, 20-88642-12, 22-88644-12, 23-88645-12, 24-88646-12, 12-89180-12, 08-88602-12, 11-88605-12, 11-88612-12, 19-88627-12, 21-88629-12, 33-88588-12, 51-88993-12, 54-88989-12, 54-88996-12, 86-89003-12, 89-89013-12, 90-89007-12, 92-89016-12, 97-89028-12, 98-89029-12, 19-88641-12, 27-88656-12, 27-88663-12, 27-88670-12, 29-88658-12, 35-88678-12, 35-88685-12, 36-88679-12, 36-88686-12, 36-88950-12, 37-88951-12, 38-88952-12, 39-88953-12, 40-88947-12, 40-88954-12, 41-88955-12, 42-88949-12, 42-88963-12, 27-88920-12, 30-88923-12, 32-88932-12, 33-88933-12, 34-88927-12, 35-88928-12, 35-88935-12, 18-88802-12, 21-88805-12, 23-88807-12, 33-88838-12, 02-89135-12, 03-89136-12, 04-89144-12, 06-89160-12, 08-89148-12, 11-89077-12, 11-89158-12, 12-89159-12, 12-89173-12, 13-89167-12, 18-89179-12, 23-89293-12, 24-89294-12, 25-89295-12, 00-89126-12, 02-89128-12, 03-89129-12, 13-88702-12, 16-88705-12, 17-88706-12, 99-89125-12, 25-88742-12, 26-88743-12, 27-88744-12, 28-88752-12, 29-88739-12, 29-88746-12, 31-88748-12, 32-88749-12, 32-88763-12, 32-88770-12, 34-88765-12, 35-88773-12, 35-88780-12, 38-88769-12, 38-88776-12, 40-88778-12, 15-88380-12, 16-88374-12, 17-88368-12, 17-88382-12, 18-88369-12, 21-88379-12, 32-88851-12, 42-88868-12, 07-89330-12, 11-89334-12, 19-88803-12, 19-88810-12, 21-88812-12, 22-88820-12, 24-88808-12, 25-88809-12, 25-88816-12, 25-88830-12, 26-88831-12, 27-88818-12, 27-88832-12, 29-88827-12, 29-88834-12, 30-88828-12, 31-88829-12, 31-88836-12, 32-88837-12, 01-89303-12, 02-89142-12, 03-89150-12, 04-89151-12, 05-89138-12, 05-89145-12, 06-89139-12, 07-89161-12, 08-89155-12, 09-89149-12, 09-89156-12, 10-89083-12, 10-89090-12, 10-89157-12, 10-89171-12, 11-89165-12, 13-89174-12, 13-89181-12, 14-89094-12, 14-89168-12, 14-89175-12, 15-89176-12, 20-89276-12, 20-89290-12, 21-89277-12, 21-89291-12, 22-89278-12, 22-89292-12, 25-89288-12, 26-89289-12, 26-89296-12, 27-89297-12, 29-89299-12, 30-88930-12, 12-88701-12, 05-89233-12, 26-88750-12, 37-88775-12, 10-89164-12, 36-88693-12, 39-88696-12, 32-88675-12, 33-88676-12, and 30-88835-12.

Root Cause

Mixed-up of materials/components

Action Taken

Animas Corporation sent an "Urgent: Insulin Pump Recall" letter dated July, 29 20011 to all affected customers. The letter included affected product and problem. They requested anyone who received the pump to send it back to Animas and a replacement pump will be sent at no charge. For additional information contact the Customer Support Department at (877) 937-7867.