Magnetic Resonance Imaging Systems, under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN
- Company
- Toshiba American Med Sys Inc
- Recall Initiated
- August 24, 2004
- Posted
- December 2, 2004
- Terminated
- September 2, 2005
- Recall Number
- Z-0273-05
- Quantity
- 39 units
- Firm Location
- Tustin, CA
Reason for Recall
Software defect causes patient identification number to be truncated.
Distribution
Nationwide.
Lot / Code Info
No specific codes were listed.
Root Cause
Other
Action Taken
Firm issued a letter notifying users of systems that when transferring image data to a DICOM server for any patient born on January 1, to verify that the Patient ID is correct on the DICOM server after the transfer. Firm will modify the software by January 2005.