Magnetic Resonance Imaging Systems, under the following names: Visart Excelart Excelart P2 Excelart P3 Excelart SPIN

Company
Toshiba American Med Sys Inc
Recall Initiated
August 24, 2004
Posted
December 2, 2004
Terminated
September 2, 2005
Recall Number
Z-0273-05
Quantity
39 units
Firm Location
Tustin, CA

Reason for Recall

Software defect causes patient identification number to be truncated.

Distribution

Nationwide.

Lot / Code Info

No specific codes were listed.

Root Cause

Other

Action Taken

Firm issued a letter notifying users of systems that when transferring image data to a DICOM server for any patient born on January 1, to verify that the Patient ID is correct on the DICOM server after the transfer. Firm will modify the software by January 2005.