Disetronic Dahedi insulin infusion pump; catalog # 8500019.
This recall has been terminated (originally issued September 23, 2003).
- Company
- Roche Diagnostics Corp.
- Recall Initiated
- July 28, 2003
- Posted
- September 23, 2003
- Terminated
- January 12, 2007
- Recall Number
- Z-1244-03
- Quantity
- 508
- Firm Location
- Indianapolis, IN
- Official Source
- View on FDA website ↗
Reason for Recall
Lack of assurance of reliability.
Distribution
United States.
Lot / Code Info
All units
Root Cause
Other
Action Taken
A recall letter dated 8/5/03 was issued to each user instructing them that they will receive a new pump and, upon receipt of the new pump, they will be required to return their Dahedi pump to the firm's representative.
More recalls by Roche Diagnostics Corp.
Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN.
Test str...
Sep 16, 2009
Roche ACCU-CHEK Performa blood glucose meters/kits, Roche Diagnostics, Indianapolis, IN.
Test str...
Sep 16, 2009
Roche/Hitachi Preciset TDM II, Roche Diagnostics, Indianapolis, IN; Catalog No. 03375781 190.
Cal...
Sep 10, 2009