Diamond Mammographic X-Ray System, type MGX-2000, model 5111771. GE Healthcare P.O. Box 414, Milwaukee, Wisconsin 53201 U.S.A.

This recall has been terminated (originally issued June 28, 2007).

Company
General Electric Med Systems Llc
Recall Initiated
November 16, 2005
Posted
June 28, 2007
Terminated
February 24, 2008
Recall Number
Z-0976-2007
Quantity
106
Firm Location
Waukesha, WI
Official Source
View on FDA website ↗

Reason for Recall

GE Healthcare has become aware of the absence of thread-locking compound associated with the C-arm fasteners of your Diamond mammographic x-ray system that may impact patient safety.

Distribution

Worldwide, including USA, Germany, France, Italy, South Africa, Finland, Norway, Sweden, Venezuela, Benelux, Japan, China, and the UK.

Lot / Code Info

Model 5111771 (foreign) 116, 117, 135, 171,175, 208, 209, 212, 215, 156, 106, 214, 211, 120, 169, 125, 119, 210, 126, 124, 139, 196, 140, 218, 121, 197, 213, 216, 170, 217, 185, 195, 152, 174, 130, 173, 138 (USA) 155, 167, 172, 176, 182, 051, 101, 118, 122, 123, 127, 128, 129, 131, 132, 133, 134, 136, 137, 141-151, 153, 154, 157-166, 177-181, 183, 184, 186-194, 198-207, 219.

Root Cause

Other

Action Taken

A Product Safety Notification, dated 05/24/07, was sent to Hospital Administrators, Managers of Radiology and Radiologists. This letter describes the safety issue, instructions to continue the use of the product and states that the device will be corrected by GE Healthcare service engineers.