Roche IGGT COBAS INTEGRA Immunoglobulin G (Turbidimetric); Catalog no. 20766631322. in vitro diagnostic.

This recall has been terminated (originally issued January 17, 2007).

Company
Roche Diagnostics Corp.
Recall Initiated
December 15, 2006
Posted
January 17, 2007
Terminated
April 23, 2008
Recall Number
Z-0356-2007
Quantity
18,905
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Falsely elevated IGGT results may be reported because the R2 reagent of the Tina-quant Gen.2 albumin assay has a carryover effect on the Immunoglobulin G (turbidimetric) assay on the COBAS INTEGRA 700 and 800 analyzers and the R1 reagent of the Tina-quant Hemoglobin A1c assay has a carryover effect on the Immunoglobulin G (turbidimetric) assay on the COBAS INTEGRA 800 analyzer.

Distribution

Nationwide.

Lot / Code Info

Lots: 66160001; exp. 12/31/06; 66395101, exp. 2/28/07; 67306401, exp. 1/31/08; 66954901, exp. 5/31/07; 67589901, exp. 4/30/08 and 68049501, exp. 10/31/08.

Root Cause

Other

Action Taken

Device correction letters dated 12/15/06 instructing users to perform an extra wash cycle between applications when using both the ALBT2 and IGGT cassettes on the COBAS INTEGRA 700 and 800 analyzers. The recall was extended to via device correction letters dated 2/28/07 instructing users to perform an extra wash cycle between applications when using both the A1C2 and IGGT cassettes on the COBAS INTEGRA 800 analyzer.