RecallDepth

IMx total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate, one bottle of 4 Methylumbelliferyl Phosphate and one bottle of Specimen Diluent

Company
Abbott Health Products, Inc.
Recall Initiated
March 24, 2005
Posted
June 16, 2005
Terminated
September 29, 2005
Recall Number
Z-0894-05
Quantity
793 kits
Firm Location
Barceloneta, PR

Reason for Recall

Abbott Diagnostics is investigating a recent increase in customer complaints for IMx calibration error code 157 and controls out of range when using IMx total B-hCG Reagent Lot 25459Q100. Testing has shown that protein aggregates are present in the specimen diluents of both IMx and the AxSYM Total B-hCG reagent kits. This is responsible for the calibration errors and shift in control values.

Distribution

US Goverment, Domestic and International distribution.

Lot / Code Info

25459Q100

Root Cause

Other

Action Taken

Abbot Health Products, Inc informed and instructed their customers via Device Recall Letter to Identify if they have used or have inventory of the affected product. Discontinue use of and destroy any remaining inventory of reagent lots listed in the Device Recall letter according to their laboratory procedures. Record the quantity of the recalled reagents currently in inventory in their facility that will be destroyed on the attached reply form and fax the form back to Abbott 1-800-777-0051. (US only) Follow their laboratory''s procedure for communicating this issue to the health care providers they serve.