IMx HAVAB Controls In Vitro Test

Company
Abbott Health Products, Inc.
Recall Initiated
March 4, 2005
Posted
March 29, 2005
Terminated
September 29, 2005
Recall Number
Z-0660-05
Quantity
373 Kits
Firm Location
Barceloneta, PR

Reason for Recall

Abbott identified through investigational studies that IMxHAVAB Controls lot 21059Q100 is generating Negative Control values outside the upper range specified in the IMxHAVAB Package Insert.

Distribution

The recalled IMx HAVAB Controls were distributed nationwide in the United Stated, including Puerto Rico and internationally were distributed to Canada, Peru, Germany, Japan, Australia, and Singapore. There are US Government accounts.

Lot / Code Info

Lot # 21059Q100 Expiration Date: May 20, 2005

Root Cause

Other

Action Taken

Abbott identified through investigational studies that IMx HAVAB Controls lot 21059Q100 is generating Negative Control values outside the upper range specified in the IMx HAVAB Package insert. If a control value is out of the specified range, the assay run is invalid and assay recalibration may be indicated. If the control values are within range as specified in the package insert, the assay run is valid and the patient results are not impacted.