Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002

Company
Toshiba American Med Sys Inc
Recall Initiated
May 23, 2005
Posted
July 13, 2005
Terminated
May 10, 2012
Recall Number
Z-0982-05
Quantity
106
Firm Location
Tustin, CA

Reason for Recall

If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values.

Distribution

Nationwide. This is an imported product.

Lot / Code Info

Software versions 6.0er001 and 002

Root Cause

Other

Action Taken

Firm issued a recall letter which provided a procedural workaround, a software patch will be installed during routine service visits.