Aplio Diagnostic Ultrasound System (SSA-700A and SSA-770A) with software version 6.0er001 and 6.0er002
- Company
- Toshiba American Med Sys Inc
- Recall Initiated
- May 23, 2005
- Posted
- July 13, 2005
- Terminated
- May 10, 2012
- Recall Number
- Z-0982-05
- Quantity
- 106
- Firm Location
- Tustin, CA
Reason for Recall
If the measurement mode is started from the Doppler mode and then the sweep speed setting is changed by the assigned user function, the measurement of the time direction results don't correspond to the actual values.
Distribution
Nationwide. This is an imported product.
Lot / Code Info
Software versions 6.0er001 and 002
Root Cause
Other
Action Taken
Firm issued a recall letter which provided a procedural workaround, a software patch will be installed during routine service visits.