Oasis-One Action Stent Introduction System with ST-2 Soehendra Tannenbaum Biliary Stent, (11.5 French size only), Guiding Catheter: 6 FR, Minimum Accessory Channel Size: 3.7mm, Wire Guide Sold Separately, Disposable Single Use Only, Quality System ISO 9001 Certified, Rx Only, Wilson-Cook Medical, GI Endoscopy, 4900 Bethania Station Road, Winston-Salem, NC 27105

Company
Wilson-Cook Medical Inc
Recall Initiated
December 22, 2003
Posted
January 14, 2004
Terminated
March 23, 2004
Recall Number
Z-0347-04
Quantity
198 units
Firm Location
Winston Salem, NC

Reason for Recall

Mislabeling; Incorrect statement, Minimum Accessory Channel Size: 3.7mm, should state, Minimum Accessory Channel Size 4.2mm

Distribution

Product was shipped to one distributor in Japan.

Lot / Code Info

Reorder number: OATS-11.5-5, Lot number: 1637695, 1651258, 1711817, 1713997, W1730838, W1752589, W1817595; Reorder number: OATS-11.5-7, Lot number: 1637696, 1651259, 1711818, 1711819, W1730839, W1738240; Reorder number: OATS-11.5-9, Lot number: 1637697, 1651260, 1702218, 1711820, 1711821, W1738241; Reorder number: OATS-11.5-12, Lot number: 1637694, 1651257, 1683866, W1730837, W1738238, W1752587; Reorder number: OATS-11.5-15 Lot number: 1692178, 1692179, 1711816, W1738239, W1752588

Root Cause

Other

Action Taken

Consignee was notified by letter sent via fax on 12/19/2003.