Exacta-Mix 2400 pharmacy compounding system.
This recall has been terminated (originally issued July 20, 2004).
- Company
- Baxa Corporation
- Recall Initiated
- November 7, 2003
- Posted
- July 20, 2004
- Terminated
- February 24, 2004
- Recall Number
- Z-0210-04
- Quantity
- 231 systems
- Firm Location
- Englewood, CO
- Official Source
- View on FDA website ↗
Reason for Recall
If device stops due to an alarm or operater pause, inlet valves will remain open, allowing some unwanted mixing of ingredients in the valve chamber..
Distribution
Nationwide. Foreign distribution to UK. No government or military consignees.
Lot / Code Info
Catalog Number EM2400, all serial numbers.
Root Cause
Other
Action Taken
All users were contacted by phone and fax on 11/7/03 with instructions to safely use the device.