Premier brand C. difficile Toxin A, diagnstic test kits, one kit per pouch, one pouch per shipping box.

Company
Meridian Bioscience Inc
Recall Initiated
December 26, 2002
Posted
January 23, 2003
Terminated
March 29, 2004
Recall Number
Z-0463-03
Quantity
1,215 kits
Firm Location
Cincinnati, OH

Reason for Recall

Due to defective packaging, the firm cannot assure product stability through the labeled expiration date.

Distribution

The kits were distributed to end-users and distributors located nationwide; and to foreign distributors located in China, Guatemala, New Zealand, Hong Kong, Mexico, Great Britain, Belgium, Italy, Brazil, Australia, Canada, Thailand, Honduras, and Japan.

Lot / Code Info

Lot #601096.180, Exp Date 27 SEP 03;  Kit Catalog #601096.

Root Cause

Other

Action Taken

The firm mailed notices to their affected customers on 12/26/2002. The consignees were instructed to visually and physically inspect the seals on each microwell pouch for gaps. If properly sealed, the use of the microwells may be continued. Test wells in pouches with defective seals are to be quarantined and destroyed. The customers are asked to report back to the firm. All test microwell plates in defective pouches will be replaced by the firm.