i-STAT CREA Cartridge, Abbott List Number: 06F10-01, 02; IVD. Manufactured by Abbott Point of Care, Inc., Princeton, NJ.

This recall has been terminated (originally issued November 24, 2010).

Company
Abbott Point Of Care Inc.
Recall Initiated
March 17, 2010
Posted
November 24, 2010
Terminated
September 22, 2011
Recall Number
Z-0456-2011
Quantity
34525 cartridges
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

The i-STAT CREA Cartridges may produce inaccurate (depressed) results at or near the end of the labeled expiry date when stored above 25 degrees for 2 weeks.

Distribution

Worldwide Distribution: Throughout USA and Canada.

Lot / Code Info

Lot numbers A09279A and A09280A, Exp March 2010

Root Cause

Other

Action Taken

Abbott Point of Care issued Urgent Product Recall Notification letters dated March 2010 to all direct customers identifying the affected product and the actions to be taken. Customers were instructed to determine if they had cartirdges that had been stored above 25 degrees Celsius, and if so, contact Abbott for replacement or credit. Abbott Technical Support can be contacted at 800-366-8020 option 1 about this recall.