RecallDepth

i-STAT EG6+ cartridge. List Number: 03P77-25.

This recall is currently active, issued September 17, 2025. It was issued by Abbott Point Of Care Inc..

Company
Abbott Point Of Care Inc.
Recall Initiated
August 21, 2025
Posted
September 17, 2025
Recall Number
Z-2585-2025
Quantity
547,250 units
Firm Location
Princeton, NJ
Official Source
View on FDA website ↗

Reason for Recall

Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.

Distribution

US Nationwide distribution.

Lot / Code Info

List Number: 03P77-25. All lot numbers.

Root Cause

No Marketing Application

Action Taken

Abbott notified consignees via email and letter on about 08/21/2025. Consignees were instructed to download, review and implement the updated i-STAT CG8+, EG7+ and EG6+ cartridge Instructions for Use (IFU), discontinue use of i-STAT CG8+, EG7+ and EG6+ cartridges for testing of capillary whole blood samples for potassium and ionized calcium, and use venous or arterial whole blood or an alternate method for testing capillary potassium and ionized calcium. Update the pH and PO2 reportable ranges on the i-STAT 1 analyzers running i-STAT CG8+, EG7+ or EG6+ cartridges. As this feature is not customizable through the i STAT 1 handheld keypad, please see the i STAT DE Quick Reference Guide - Customizing Reportable Ranges (ART 770547-00 Revision A) for more information. To use the Custom Reportable Range feature, you must have the i-STAT 1 handheld and i STAT DE version 2.3 or higher. Changes to the reportable range may require a change to the Laboratory Information System (LIS) interface. Additionally, consignees were requested to confirm receipt and understanding of this communication by responding to the business reply card included with the letter. If consignees have forwarded any i-STAT CG8+, EG7+ or EG6+ cartridges to another facility, they were requested to notify them.

More recalls by Abbott Point Of Care Inc.

View all recalls by this company →