Bausch & Lomb Stellaris Vision Enhancement System.
This recall has been terminated (originally issued May 26, 2010).
- Company
- Bausch & Lomb Inc
- Recall Initiated
- January 15, 2010
- Posted
- May 26, 2010
- Terminated
- May 17, 2013
- Recall Number
- Z-1664-2010
- Quantity
- Quantity never provided by the firm
- Firm Location
- Rochester, NY
- Official Source
- View on FDA website ↗
Reason for Recall
The AC power cords used on these products are subject to an FDA Alert -- Safety Investigation of Certain Medical Device Power Cords: Initial Communication, Date Issued: October 19, 2009.
Distribution
Worldwide distribution: United States & Puerto Rico; Venezuela, India, Trinidad and Tobago, Colombia, Sri Lanka, Hong Kong, Vietnam, and Peru.
Lot / Code Info
All devices manufactured since 2004
Root Cause
Nonconforming Material/Component
Action Taken
A "URGENT: POWER CORD RECALL" letter and copies of the FDA Alert dated Oct. 19, 2009 were sent to all customers on or about Jan. 15, 2010, by Stericycle, a recall notification service. The letter describes the product, problem and actions to be taken by the customers as well as FDA's recommendations and actions. The customers were made aware of the FDA Alert concerning the AC power cords and notified that these power cords were used in the devices. They were requested to inspect the power cords for damage and notify the recalling firm if damage is noted or if they have experienced sparking, charring, or firing from the power cords and/or cracking or failure of the prongs. In the absence of those conditions, customers were given the option of returning the power cord for replacement and/or during the regularly scheduled servicing of the device. If you would like a new power cord, please contact Customer Service at 1-800-338-2020 for a free replacement.