RecallDepth

Bausch & Lomb Stellaris Basic Vacuum Phaco Pack, Ref: BL5111, packaged 6 units/case, RX, Sterile. Bausch & Lomb Incorporated, Rochester, NY For use with the Bausch & Lomb Stellaris Vision Enhancement System for the phacoemulsification of an opacified crystalline lens during anterior segment surgery

This recall has been terminated (originally issued August 30, 2011).

Company
Bausch & Lomb Inc
Recall Initiated
July 27, 2011
Posted
August 30, 2011
Terminated
June 21, 2013
Recall Number
Z-3088-2011
Quantity
229/6-unit cases
Firm Location
Rochester, NY
Official Source
View on FDA website ↗

Reason for Recall

The aspiration channel on the affected vacuum cassettes may be blocked due to a molding malfunction

Distribution

Nationwide Distribution.

Lot / Code Info

Model # BL5111, Lot #U6481,  Exp. 11/2012.

Root Cause

Nonconforming Material/Component

Action Taken

Bausch & Lomb Inc. notified all affected customers by direct telephone contact on July 27, 2011 and a follow-up FAX or e-mail dated July 27, 2011 The letter included; affected product, problem, and requests customers return the product. An acknowledgement form was included to complete and return with the product. For question on this recall contact Bausch & Lomb at (949) 521-7895.

More recalls by Bausch & Lomb Inc

View all recalls by this company →