RecallDepth

Spine Frontier Inspan compressors Part Number: 11-60004. Inspan compressors provide pressure to drive spikes of Inspan implants into the spinous process bone, resulting in fixation of the implant.

This recall has been terminated (originally issued November 21, 2012).

Company
Spinefrontier, Inc.
Recall Initiated
August 28, 2012
Posted
November 21, 2012
Terminated
June 11, 2015
Recall Number
Z-0412-2013
Quantity
17 units
Firm Location
Beverly, MA
Official Source
View on FDA website ↗

Reason for Recall

Inspan Compressors may break at weld causing a loss of compression

Distribution

Worldwide Distribution-USA (nationwide) including PR and the states of FL, MA, MO, OK, and TX, and the country of Mexico.

Lot / Code Info

Lot Numbers: SFP02323, SFP02323R, and SFP02323RR

Root Cause

Device Design

Action Taken

Spine Frontier issued e-mail notices to customers on May 14, 2012 advising of the potential for breakage. Units were replaced with weld free design.