RecallDepth

lnline Persuader- PedFuse Pedicle Screw System; Part Number: 11-80164

This recall has been terminated (originally issued January 14, 2015).

Company
Spinefrontier, Inc.
Recall Initiated
December 19, 2014
Posted
January 14, 2015
Terminated
January 11, 2016
Recall Number
Z-0977-2015
Quantity
4 devices
Firm Location
Beverly, MA
Official Source
View on FDA website ↗

Reason for Recall

A crack may develop in the handle to the Inline Persuader, which may progress if the user continues to use the affected instrument.

Distribution

Nationwide Distribution including TX, OH, VA, and Washington, DC.

Lot / Code Info

Lot/Serial Numbers: 06440-0000, 06441-0000, AG24

Root Cause

Device Design

Action Taken

Spinefrontier sent an Urgent Voluntary Recall Notification letter dated December 16, 2015, to all affected customers. The notice informed customers of the recall and requests that they initiate the immediate return of the parts listed. In addition, SpineFrontier is requiring confirmation via recall acknowledgement form that consignees have received this notice and that they will initiated the immediate return of the parts. Customers can contact SpineFrontier at 978-232-3990 Monday through Friday 8:30AM to 5:30PM EST with any questions.

More recalls by Spinefrontier, Inc.

View all recalls by this company →