Volcano brand Intravascular Ultrasound, Volcano s5/s5i Family Software Version 3.2.1 or 3.2.2, Models/Part Numbers: 435-0607.01, 435-0607.04, 435-0602.01, 435-0602.02; Product is manufactured and distributed by Volcano Corporation, Rancho Cordova, CA. Medical device for use in imaging of histology.

Company
Volcano Corporation
Recall Initiated
August 24, 2012
Posted
September 10, 2012
Terminated
November 25, 2014
Recall Number
Z-2345-2012
Quantity
1020 units
Firm Location
Rancho Cordova, CA

Reason for Recall

The display on the system has the potential to change. In house testing eventually identified the unique sequence of events that must occur in order for this issue to manifest itself.

Distribution

Worldwide Distribution.

Lot / Code Info

All units of these models. Models/Part Numbers: 435-0607.01, 435-0607.04, 435-0602.01, 435-0602.02

Root Cause

Software design

Action Taken

Volcano Corporation sent an "IMPORTANT FIELD CORRECTIVE ACTION" letter dated August 24, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Contact your local Volcano sales representative or Volcano Customer Service for questions regarding this recall.