Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

This recall has been terminated (originally issued July 20, 2012).

Company
Leica Microsystems, Inc.
Recall Initiated
June 4, 2012
Posted
July 20, 2012
Terminated
October 29, 2012
Recall Number
Z-2053-2012
Quantity
39 Units - USA , 4 Units - Canada
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.

Distribution

Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, West Virginia, including Puerto Rico, and the countriy of Canada.

Lot / Code Info

Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001

Root Cause

Software design

Action Taken

Leica Microsystems sent an Urgent Medical Device letter dated June 4, 2012 to the affected customers. The letter described the product, problem and actions to be taken. The customers were informed that the Leica surgical microscope M525 F50 required a software upgrade. Customers were instructed to NOT USE a remote video adapter or motorized video adapter until the required software upgrade has been installed by their Leica Microsystems representative. The customers were requested to complete the attached Acknowledgement Form, acknowledging receipt and understanding of the letter, listing the serial numbers of the units at their facilities and to fax it back to Leica Microsystems at (847) 236-3747.