Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Company
Aesculap Inc
Recall Initiated
December 17, 2014
Posted
January 26, 2015
Terminated
December 13, 2016
Recall Number
Z-1012-2015
Quantity
772
Firm Location
Center Valley, PA

Reason for Recall

Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

Distribution

Worldwide Distribution - US Nationwide

Lot / Code Info

material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T

Root Cause

Labeling False and Misleading

Action Taken

Aesculap sent an Important Correction and Removal and Field Safety Notification letter, dated December 17, 2014, to all affected customers. The letter identiifed the affected product, problem, and actions to be taken. Customers were requested to remove from inventory and contact Aesculap regarding return of the affected devices and acknowledgement form.