MC-500 Multicolor Laser Photocoagulator; Distributed by NIDEK Inc. Manufactured by NIDEK CO., Ltd, Japan. For use in ophthalmic and retinal procedures.

This recall has been terminated (originally issued July 30, 2014).

Company
Nidek Inc
Recall Initiated
June 9, 2014
Posted
July 30, 2014
Terminated
November 24, 2014
Recall Number
Z-2132-2014
Quantity
29 units (26 already re-labeled)
Firm Location
Fremont, CA
Official Source
View on FDA website ↗

Reason for Recall

Laser aperture label was not affixed to device prior to shipment in the US.

Distribution

Nationwide Distribution including IL NE NY, NH, CD, MI, TX, CA, PA, FL, KS, UT, MN, WI,. MO. and OR.

Lot / Code Info

Part number: 173530U20 Serial numbers: 50010, 50127, 50136, 50277, 50310, 50311, 50330, 50331, 50357. 50358, 50359, 50382, 50383, 50408. 50417, 50418, 50426, 50439, 50445, 50446, 50471, 50514, 50552, 50558, 50573, 50600, 50610, 50611, 50651.

Root Cause

Error in labeling

Action Taken

Nidek sent a Recall/Field Correction letter dated June 9, 2014, to all affected customers. The letter advised customers of the issues and informed customers that Nidek Representatives will visit each location to affix the proper labeling to the device. Customers with questions were instructed to contact Nidek Customer Service at 1-800-223-9044 M-F 8 am to 5 pm PST. For questions regarding this recall call 510-353-7785.