BG3515-5 US 5mL BioGlue Surgical Adhesive. BioGlue Surgical Adhesive (BioGlue) is indicated for use as an adjunct to standard methods of achieving hemostasis (such as sutures and staples) in adult patients in open surgical repair of of large vessels (such as aorta, femoral, and carotid arteries).
- Company
- Cryolife Inc
- Recall Initiated
- April 25, 2014
- Posted
- May 30, 2014
- Terminated
- November 10, 2014
- Recall Number
- Z-1674-2014
- Quantity
- 6190 units
- Firm Location
- Kennesaw, GA
Reason for Recall
Serum albumin component monomer failed to meet internally established end of shelf-life specification.
Distribution
Distributed Nationwide including Puerto Rico and the states of AL, AK,AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, MT, NV, NH, NJ, NM , NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WV, and WI.
Lot / Code Info
12MUV034, 12MUX035, 12MUV069, 12MUV037, 12MUV038, 12MUV040, 12MUV041, 12MUV042, 12MUV043, 12MUV044, 12MUV045, 12MUV046, 12MUV047, 12MUV048, 12MUV049, 12MUV050, 12MUV051, 12MUV052, 12MUV053, 12MUV054, 12MUV055, 12MUV056, 12MUV057, 12MUV058, 12MUV059, 12MUV060, 12MUV061, 12MUV062, 12MUV063, 12MUV064, 12MUV065, 12MUV067, 12MUV068, 12MUV069, 12MUV070, 12MUV071, 12MUV072, 12MUV073, 12MUV074, 12MUV075, 12MUV076, 13MUV001, 13MUV002, 13MUV003, 13MUV004,
Root Cause
Process control
Action Taken
All consignees were notified of the voluntary actions by issuance of Field Correction/Removal Notice on April 24, 2014. This notices states which lot numbers are to be returned, and which lot numbers can be relabeled with a corrected expiration date.