RecallDepth

On-X Aortic Heart Valve with Anatomic Sewing Ring Product Usage: The On-X Prosthetic Heart Valve is indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic heart valves in the aortic position. It is a bileaflet mechanical heart valve, which consists of an orifice housing and two leaflets. The On-X Prosthetic Heart Valve is available in 3 aortic configurations. All aortic configurations are available in sizes 19, 21, 23, 25, and 27/29 mm. Aortic valves, size 19 mm through 25 mm, are designed for intrasupra-annular sewing ring position, while the valve size 27/29 mm is designed for intra-annular sewing ring position. All mitral valve sizes are designed for the supra-annular sewing ring position.

Company
Cryolife Inc
Recall Initiated
January 17, 2019
Terminated
June 9, 2020
Recall Number
Z-1073-2019
Quantity
2 devices
Firm Location
Kennesaw, GA

Reason for Recall

On-X Valve was mislabeled with the incorrect serial number.

Distribution

Worldwide Distribution - US Nationwide - Georgia and Germany.

Lot / Code Info

Model/Catalog Number ONXANE - 23, Lot number 6827216, Expiration Date 11/19/2024

Root Cause

Employee error

Action Taken

The affected customer was initially notified via email on 01/16/2019. A follow up notification explaining the reason for recall was provided via email on about 01/24/2019. Instructions included to quarantine and return the affected device.

More recalls by Cryolife Inc

View all recalls by this company →