MXR-35 Diagnostic X-ray High Voltage Generator manufactured by DRGEM Corporation. x-ray generator

Company
Fujifilm Medical System Usa, Inc.
Recall Initiated
June 29, 2012
Posted
May 21, 2014
Terminated
June 1, 2018
Recall Number
Z-1642-2014
Quantity
14
Firm Location
Stamford, CT

Reason for Recall

The MXR-35 X-ray Generator's labeling did not include the certification statement as required by 21 CFR Part 10 10.2.

Distribution

Worldwide Distribution - US Distribution including the states of TX and KY and Internationally to China.

Lot / Code Info

SN: MXA1170108, MXA1270262

Root Cause

Radiation Control for Health and Safety Act

Action Taken

Fujifilm will replace the violative labels with replacement labels provided by DRGEM, to bring the X-ray generator into compliance. You stated that you expect for all nine installed units to be corrected by the end of May 2014. The seven units in the Stamford, CT warehouse will be corrected when the generators are delivered to customer sites. Under 21 CFR 1003.31(c)(2), this exemption is granted subject to the following conditions: 1. You must correct the missing certification statement for all units shipped into the US within the timeline you specified, by the end of May 2014. 2. You must ensure that all future units that will be delivered and installed in the US will be in compliance with the labeling requirement, per 21 CFR 1010.2.