Genital Protector Shield used with Fuji Computed Radiography (FCR) system models 5501D and XU/D1. Model: GP344
- Recall Initiated
- September 30, 2004
- Posted
- October 28, 2004
- Terminated
- December 22, 2004
- Recall Number
- Z-0055-05
- Quantity
- 16
- Firm Location
- Stamford, CT
Reason for Recall
Genital Protector Shield used with Fuji Computed Radiography (FCR) system models 5501D and XU/D1. Model: GP344 may detach from the mounting block and fall during use.
Distribution
CO, CT, NY, MD, MI, MN, NV, UT
Lot / Code Info
All lots
Root Cause
Other
Action Taken
Fuji are notifying users through Fuji service representatives who are hand delivering the notfication letter dated 9/30/04 and installing the mandatory hardware upgrade.