Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (NCL-L-GSTpi-438), a liquid tissue culture supernatant. For in vitro diagnostic use.

This recall has been terminated (originally issued June 2, 2014).

Company
Leica Microsystems, Inc.
Recall Initiated
February 26, 2014
Posted
June 2, 2014
Terminated
August 14, 2017
Recall Number
Z-1698-2014
Quantity
33 units
Firm Location
Buffalo Grove, IL
Official Source
View on FDA website ↗

Reason for Recall

Novocastra Liquid Mouse Monoclonal Antibody Glutathione S-Transferase pi (Product Code: NCL-L-GSTpi-438) does not function as intended up to the expiry date on the product labeling. This was detected through an ongoing stability program. There is a linkage between the age of the product and staining intensity.

Distribution

Nationwide Distribution including AZ, CA, FL, and NY.

Lot / Code Info

Product Code: NCL-L-GSTpi-438; Lot Numbers and Expiration Dates: 6018777, Expires 2014-05; 6023260, Expires 2014-12; 6026297, Expires 2015-05; 6026980, Expires 2015-06

Root Cause

Incorrect or no expiration date

Action Taken

Leica Biosystems sent a Medical Device Recall Notification letter dated February 26, 2014 to all direct accounts. The letter advised that clinicians may want to review results obtained from the affected products. The letter also included instructions for customers to: 1) appropriately destroy any unused or partially used NCL-L-GSTpi-438 (lots 6018777, 6023260,6026297, 6026980) reagent or confirm that they are not in stock; 2) complete and return the attached Medical Device Recall Notification Acknowledgement Form to 847-236-3747; and, 3) contact the local Leica representative immediately if replacement is required. Customers with questions or concerns about this recall can contact Leica via e-mail at LMGRA@leica-microsystems.com. For questions regarding this recall call 847-405-5413.