RecallDepth

Ziehm Vision RFD C-Arm Interventional Fluoroscopic X-Ray System. Mobile Fluoroscopic Imaging

This recall has been terminated (originally issued December 24, 2013).

Company
Ziehm Imaging Inc
Recall Initiated
November 21, 2013
Posted
December 24, 2013
Terminated
June 23, 2014
Recall Number
Z-0568-2014
Quantity
27 devices
Firm Location
Orlando, FL
Official Source
View on FDA website ↗

Reason for Recall

Engineering change in 2006 resulted in a minor non-conformity of their devices. Specifically the emergency switch was changed from a standard component to just an option on the devices.

Distribution

USA (Nationwide) Distribution including the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MT, MA, MD, MI, MN, MN, MO,MS, NC ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Hawaii.

Lot / Code Info

20006 20025 20142 20185 20216 20231 20015 20026 20151 20186 20217 20236 20016 20027 20153 20189 20225 20018 20031 20166 20198 20227 20019 20139 20184 20199 20228

Root Cause

Device Design

Action Taken

Ziehm Imaging sent an "URGENT RECALL NOTICE" letter dated November 22, 2013, to all affected customers. The letter identified the products the problem and the action needed to be taken by the customers. Customers were advised and instructed to: Your Model and Serial number can be found on the product label attached to the side of the mobile C-arm Stand. Our records indicate that your facility has one or more of the products listed above that are being used in your facility. Recall of the product was initiated because according to our manufacturing documentation, the referenced device installed at your facility was not equipped with an emergency stopping device for vertical travel of the c-arm imaging assembly. To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device identification and label. The issue: According to our manufacturing documentation, the referenced device installed at your facility was not equipped with an emergency stopping device. To fully comply with applicable standards the referenced device must be retrofitted with an emergency stopping device. The manufacturer Ziehm Imaging GmbH advised customers they may continue to operate the affected device with caution until the correction has been carried out on-site. During this time Ziehm Imaging GmbH advises the operator to be aware of the following safety measures which may resolve the possibility of potentially hazardous situations by: " Operating the power-off switch located either on the C-arm Stand top panel or on the monitor cart " Removing the power cable from the wall outlet or, " Activate the push-buttons for reverse vertical movement and keeping the pushbutton pressed until patient /operator is no longer in the risk zone or the device stops Please alert all staff members working with the device to these potential hazardous situations, and familiarize them with all possibilities to resolve the hazardous situation

More recalls by Ziehm Imaging Inc

View all recalls by this company →