Ziehm Vision R, Mobile C-arm for Mobile Fluoroscopic Imaging

This recall has been terminated (originally issued April 15, 2016).

Company
Ziehm Imaging Inc
Recall Initiated
April 15, 2016
Terminated
July 18, 2019
Recall Number
Z-2642-2016
Quantity
1 device
Firm Location
Orlando, FL
Official Source
View on FDA website ↗

Reason for Recall

Malfunction of the radiation-warning timer and the 3D-specific interval warning. No 5-minute timer dialogs or beeper will occur until the unit is rebooted.

Distribution

US Distribution including Puerto Rico and to the states of :FL, GA, VA,MA. ME, MI, NY, NJ, PA, RI, IL, TX, AZ, CA, NV, UT, and WA.

Lot / Code Info

Serial # 10841.

Root Cause

Device Design

Action Taken

Ziehm Imaging, Inc planned action to bring these defects into compliance: 1. You will contact customers and initiate a software update to correct the defects. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call (866) 949-4346.