Artiste MV Digital Linear Accelerator System Equipped with 160 MCL Accessory Holder, Manufactured by Siemens Medical Solutions, Oncology Care System, Concord, CA Medical charged-particle radiation therapy system.

This recall has been terminated (originally issued January 30, 2009).

Company
Siemens Medical Solutions Usa, Inc
Recall Initiated
December 20, 2008
Posted
January 30, 2009
Terminated
December 9, 2010
Recall Number
Z-0949-2009
Quantity
43 units
Firm Location
Concord, CA
Official Source
View on FDA website ↗

Reason for Recall

The range of tolerance is to high for the accessory holder, therefore, Wedges and other accessories do not latch to the locking mechanism correctly.

Distribution

Nationwide and Worldwide

Lot / Code Info

All units of Part Number: 8139789

Root Cause

Device Design

Action Taken

On 12/20/08, the firm issued letters via certified mail to direct consignees, informing them of the affected product and providing instructions on the recall. Contact Siemens Medical Solutions USA, Inc. at 1-925-246-8407 for additional information.