High voltage generator XTP-8100G and XTPG-8100G for the Infinix-i series interventional angiography system.

This recall has been terminated (originally issued November 24, 2005).

Company
Toshiba American Med Sys Inc
Recall Initiated
September 21, 2005
Posted
November 24, 2005
Terminated
July 12, 2007
Recall Number
Z-0198-06
Quantity
30
Firm Location
Tustin, CA
Official Source
View on FDA website ↗

Reason for Recall

It was found during a study, arcing might occur in the transformer of the high voltage generator, XTP-8100G and XTBP-8100G displaying an error message and interrupting fluoroscopy.

Distribution

CA, CO, KS, IL, LA, IN, ME, MI, NC, NJ, NY, OH, PA, TX

Lot / Code Info

model numbers: XTP-8100G AND XTBP-8100G

Root Cause

Other

Action Taken

Consignees notification that a local TAMS service representative will schedule the implementation of the replacement transformer.